SE LogoOtO Photonics' SmartEngine (SE) Series of compact Czerny-Turner spectrometers can be specified with a huge range of configurations based on a choice of:-

 

  • 8 standard sensors
  • 6 standard spectral response ranges (gratings)
  • 6 standard slit sizes

 

For volume opportunities with OEM customers OtO also develop bespoke configurations with custom gratings and slits along with non-standard sensors, optics and interface connections.

 

SmartEngine's standard model numbers and associated sensors are:-

 

Model Sensor Characteristic Dynamic
Range
SNR
SE2020 Sony ILX511B
UV-enhanced front illuminated CCD
Full 180-1100nm 1300 250
SE2030 Hamamatsu S11639
CMOS with fastest exposure
Full 180-1100nm
Shortest exposure
1600 330
SE2040 Sony ILX554B
Front illuminated CCD
200-1025nm
Excellent cost:performance
1800 200
SE2050
SE2052
Hamamatsu S11510
NIR-enhanced back-thinned CCD
790-1010nm
High SNR @ NIR
5000 500
SE2060 Hamamatsu S10420
UV-enhanced back-thinned CCD
180-1025nm
High SNR @ UV
5000 500
SE2070 Toshiba 1304
High pixel-resolution CCC
200-1025nm
High resolution
1100 400
SE2080 Hamamatsu S13496
CMOS with fast exposure
200-1025nm
Short exposure (0.4ms)
High resolution
936 350
SE2090 Hamamatsu S11071
Back-thinned CCD with fast exposure
180-1100nm
Fast exposure (2ms)
2184 500

 

These standard models can be specified with a variety of standard spectral response ranges based on 6 standard grating designs:-

 

DUVN: 200-1025nm, 500 lines blazed at 300nm

FUV: 180-850nm, 600 lines blazed at 300nm

FUVN: 180-1100nm, 300 lines blazed at 300nm - normally used with SE2020

FUVN2: 180-1100nm, dual zone, 300 lines blazed at 250nm/580nm, 70/30 ratio - normally used with SE2060

VNIR: 350-1020nm, 600 lines blazed at 500nm

NIR1: 790-1010nm, 1200 lines blazed at 850nm

 

Standard combinations of sensor and grating (spectral range) are:-

 

Model Spectral Response Range (nm)
DUVN FUV FUVN FUVN2 VNIR NIR1
200-1025 180-850 180-1100 180-1100 350-1020 790-1010
SE2020 * * * * *
SE2030 * * * * *
SE2040 * *
SE2050
SE2052
  *
SE2060   * * *
SE2070 * * *
SE2080 * * *
SE2090   * * *

 

Finally - users can select from 6 standard slit sizes. In combination with the 6 standard gratings (spectral ranges) and other non-standard gratings described below it is possible to derive a specification for spectral range and resolution:-

 

Groove
Density
(g/mm)
Optimum
Efficiency
Wavelength

Bandwidth

Selectable
Band
(nm)
Resolution (nm)
10µm 25µm 50µm 100µm 200µm 300µm
2400 400nm 100nm 200-600 0.2 0.25 0.4 0.7 1.2 1.8
1800 250nm 150nm
UV 200nm
190-380 0.3 0.4 0.6 1.0 1.8 2.6
1200 200nm
300nm
500nm
600nm
750nm
850nm (NIR1)
220nm 180-1010 0.4 0.5 0.8 1.3 2.3 3.3
1000 250nm
900nm
300nm 180-1100 0.5 0.7 1.1 1.9 4.0 6.5
900 500nm 400nm 180-1100 0.6 0.8 1.3 2.3 4.6 7.1
600 300nm (FUV)
400nm
500nm (VNIR)
800nm
1000nm
670nm 180-1100 0.9 1.2 1.9 3.2 6.0 8.7
500 300nm (DUVN)
565nm
770nm
825nm 180-1100 1.0 1.3 2.0 3.5 6.5 9.5
400 250nm - 180-1100 1.1 1.5 2.2 3.8 6.9 9.5
300 300nm (FUVN)
280/580nm (FUVN2)
920nm 180-1100 1.5 2.0 3.0 5.3 10.5 -

 

By combining the choice of grating, slit and sensor it is possible to derive an optimum configuration for your application - OtO also have recommendations for specific applications.

 